Results

What the three levels of the method found. The parser, the calibration and the start of a capture are those of this release. See Hardware tests for what changed after the captures.

Level Result Run on
The parser, 600 samples Every current equal to three decimals, every digital byte identical The current code
Captures in Ampere meter mode Mean current and mean charge of the marked regions within 0.11 % of the app's 2026-11-21
Captures in Source meter mode Mean current and mean charge of the marked regions within 0.04 % of the app's 2026-11-21
The 0.2 µA floor No value between 0 and 0.2 µA, in either tool 2026-11-21
A .ppk2 file opened in the app Loads; average, maximum and charge of a 10 s window equal to those of the CSV 2026-11-21
Ten hardware tests All pass 2026-11-21 in both modes, 2026-11-22 in Source meter mode

Parser

On the 600 recorded samples, this package and the app's parser give the same current to three decimals for every sample, and the same digital inputs. The stream crosses all five measurement ranges, so the calibration of each range and the spike filter are part of what agrees.

Captures

The device under test (DUT) is an nRF54LM20 DK that runs the bench firmware: D0 toggles at 5 Hz and D1 at 0.5 Hz, and the CPU is busy while D0 is high.

The bench was captured for 30 s by the app and by ppk2 capture, within minutes of each other and without being touched in between.

Item Value
Date 2026-11-21, between 19:31 and 19:40
PPK2 One unit, firmware 1.2.4 (db16a94), the version the app installs. Its calibration text is in the evidence folder
App Power Profiler 4.4.1, in nRF Connect for Desktop 5.4.1
This package Commit 39675d7, with Python 3.13.13 and pyserial 3.5
Host macOS 26.5.1
DUT nRF54LM20 DK with the bench firmware
Captures 30 s each at 100,000 samples per second. Those of ppk2 capture start at power-on, the app's about 0.7 s before it

How to read the tables:

  • The current of a region is its mean current, and its charge is that current added up over its width.
  • Each is given as the mean ± the standard deviation over the regions.
  • A difference is that of the means, as a share of the app's, computed before rounding.
  • The KS statistic compares the distributions of the current over the same 26 s, starting 2 s after power-on.

Ampere meter mode, 3.3 V

The app's capture was taken between two of ppk2 capture.

App (19:34) ppk2 capture (19:31) ppk2 capture (19:35)
D0 regions 143 147 147
Current of a D0 region 1,380.4 ± 1.8 µA 1,381.6 ± 2.0 µA 1,380.9 ± 1.6 µA
Charge of a D0 region 138.02 ± 0.18 µC 138.14 ± 0.20 µC 138.07 ± 0.16 µC
D1 regions 14 14 14
Current of a D1 region 699.3 ± 2.6 µA 699.9 ± 3.0 µA 700.1 ± 2.6 µA
Charge of a D1 region 699.3 ± 2.6 µC 699.9 ± 3.0 µC 700.1 ± 2.6 µC
Idle median 1.40 µA 1.40 µA 1.40 µA
Difference from the app, D0 current and charge 0.09 % 0.04 %
Difference from the app, D1 current and charge 0.09 % 0.11 %
KS statistic against the app 0.0037 0.0032

A D1 region lasts one second, so its charge in µC is, to rounding, its current in µA.

The two captures of ppk2 capture differ from each other by 0.05 % on the D0 current. So the difference from the app is of the size of the bench's own drift over four minutes.

Both captures in Ampere meter mode, overlaid from power-on, and the distribution of the current

Source meter mode, 3.3 V

Here both captures of ppk2 capture follow the app's.

App (19:38) ppk2 capture (19:38) ppk2 capture (19:39)
D0 regions 143 147 147
Current of a D0 region 1,385.92 ± 0.82 µA 1,385.4 ± 1.0 µA 1,385.64 ± 0.79 µA
Charge of a D0 region 138.570 ± 0.087 µC 138.52 ± 0.11 µC 138.541 ± 0.086 µC
D1 regions 14 14 14
Current of a D1 region 700.9 ± 2.8 µA 700.7 ± 1.8 µA 700.8 ± 2.2 µA
Charge of a D1 region 700.9 ± 2.8 µC 700.7 ± 1.8 µC 700.9 ± 2.2 µC
Idle median 1.35 µA 1.35 µA 1.35 µA
Difference from the app, D0 current and charge 0.04 % 0.02 %
Difference from the app, D1 current and charge 0.03 % 0.00 %
KS statistic against the app 0.0034 0.0056

Both captures in Source meter mode, overlaid from power-on, and the distribution of the current

The 0.2 µA floor

The app records any current below 0.2 µA as zero, and so must ppk2 capture. Over the whole of each capture, the unpowered start included:

App ppk2 capture 1 ppk2 capture 2
Ampere meter: values between 0 and 0.2 µA 0 0 0
Ampere meter: negative values 0 0 0
Ampere meter: lowest value above zero 0.204 µA 0.201 µA 0.201 µA
Source meter: values between 0 and 0.2 µA 0 0 0
Source meter: negative values 0 0 0
Source meter: lowest value above zero 0.215 µA 0.215 µA 0.215 µA

Session file

A 30 s capture in Source meter mode at 3.3 V was converted with ppk2 convert and loaded in the app. For the last 10 s of the trace:

The app, from the .ppk2 file Computed from the CSV
Average 699.32 µA 699.32 µA
Maximum 171.80 mA 171.80 mA
Charge 6.99 mC 6.993 mC
D0 pulses 50 50
D1 pulses 5 5

The trace ends at 30.00319 s in both. This checks that the app reads what ppk2 writes. The two files are not in the repository: the CSV is 123 MB.

Hardware tests

All ten pass, at 3.3 V on an nRF54LM20 DK. They were run twice:

  • on 2026-11-21, in both modes, on the code of the captures above;
  • on 2026-11-22, in Source meter mode, on the code of this release.

Between the two runs, the way a capture ends and the handling of errors changed: a capture now stops on its count of samples, and fails when they do not all arrive. The parser, the calibration and the start of a capture, which the captures above check, did not change.

One more check was made by hand: a running capture was frozen for 3 s, which makes the computer lose samples. Three such captures recovered, with 18 to 34 samples marked lost and normal currents afterwards. See the protocol page for how the parser finds its place again.

Differences from the app

On purpose, and outside what the comparisons above exercise:

  • Lost samples. The two parsers differ in five places, all on a stream with lost or damaged data. On such a stream this package emits each held sample once, counts a lost gap in full, and keeps the samples in order. The app emits a held sample on arrival and again when it releases it. It counts the gap one sample short and places the samples it held before the gap. See the protocol page for the five points, with their line references. Both tools record a lost sample as 0.2 µA, and no capture above holds such a sample: the lowest value above zero is 0.201 µA.
  • Where a capture starts. In the app you start the capture and enable power by hand. ppk2 capture does both, and its file begins about 10 ms before the power comes up.

Limitations

  • The bench: one PPK2 and one dev kit, on a Mac. The package runs on Linux and Windows too, where its unit tests pass, but no capture was compared with the app there.
  • The reference is the app. Agreement with it says nothing about the absolute accuracy of the PPK2, which Nordic gives as better than ±20 %.
  • Each capture lasts 30 s. Longer captures were not compared.

Reproducing the results

The captures and the scripts are in validation/evidence/bench_dut/, the recording and its reference in tests/fixtures/. From a checkout:

# the parser against the stored reference
just test

# the six captures, from the branch `captures` of the repository (42 MB)
just captures

# the tables above, from those captures
uv run --group validation python validation/evidence/bench_dut/summary.py

# the plots above
just plots

# current and charge per marked region of one capture
python validation/regions.py validation/evidence/bench_dut/captures/source_app.csv.xz

# two captures of your own bench
uv run --group validation python -m validation.compare_cli app.csv ours.csv --skip-initial-s 2